Accessories for Exposure Systems

VITROCELL®Vacuum Flow Calibration

Vacuum flow adjustment for aerosol delivery to cell cultures

The VITROCELL® flow calibration valves regulate the flow of the aerosol through the modules in the ml range.

The valves are located between the vacuum pump and the aerosol exposure top. Mass flow meters enable accurate valve read-outs. The consistent and adjustable flow rate ranges from 0…20 ml/min. Alternative flow rate ranges are available upon request.

The valves enable fast control of the flow rates prior to the experiment. They are manufactured from Teflon/stainless steel and can be easily cleaned. If the aerosol contains particles, we recommend installing disposable microfiber filters in front of the valves.

flow-calibration-up.png

VITROCELL®Mass Flow Meters

For clean and dilution air flow measurement

Mass flow meters are required to adjust the vacuum flow by means of vacuum calibration valves and to accurately set dilution and aerosol generator air flows.

They are available in ml/min to l/min. ranges.

mass-flow-up.png

VITROCELL®Mass Flow Controller

Specifically designed to precisely regulate dilution and aerosol generator airflow.

Mass Flow Controller are available for various ml/min ranges.

Mass-Flow-Controller-various-ranges.jpg.webp.jpeg

More about Mass Flow Controller

VITROCELL® Mass Flow Control System
vc1-7-mode-1-640x427.jpg.jpeg

Direct connection of draw-actuated cigarettes to the syringe drives.

Mode 2 for e-cigarettes
Mode-2-for-e-cigarettes-640x427.jpg.jpeg

7 individual vapestarter units with button activators for tank products are connected to the syringe drives.

VITROCELL®Clean Air Purification Station

Filter system to improve the quality of compressed air

The VITROCELL® Clean Air Purification Station is recommended when the air quality of a central air supply system or of a lab compressor contains oil or particles which may influence the test results. 

An indication of impurities in the system and thus the need for an improvement of the air quality is evident if the cell viability drops substantially in the clean air exposure modules.

Clean-Air-Purification-Station-detail.jpg.webp

PowerVent components

Clean-Air-Purification-Station-front-back-sub1.png

VITROCELL® Water Bath

For up to 3 modules

A perfect solution for heating up to 3 exposure modules.

VITROCELL® cultivation and exposure modules can be operated without an incubator. The modules can be heated to maintain a media temperature of 37º C (99º F). For the VITROCELL® 6/6, 12/12, 24/24 and 24/48 systems, heating is controlled electrically. All other modules are heated by warm water, either by water circuit or by a heating plate for optimal heat transferal to the module. This water bath is suitable for up to 3 exposure modules.

Ensuring accuracy in temperature control with a modern state-of-the-art design. These units provide best-in-class economical performance while remaining user-friendly by reducing unnecessary functions and focusing on reliability. Engineered from highest-quality components and materials, such as stainless steel for the baths, durable heating thermostats with a long operation life are guaranteed.

water-bath-up.png

Features

  • For up to 3 modules

  • 3-Key operation with LED-Display

  • LED temperature display

  • Temperature Range 25° C to 100° C

  • Prolonged operating life

  • Dimensions: 147 x 307 x 330 mm (W x D x H)

  • Delivered with cover plate and tubing connectors

  • Max. flow 20 l/min

  • Max. pump pressure 0,2 bar

  • Bath volume max. 6 Liters

  • Filling volume min. 4.9 Liters

VITROCELL® High Performance Water Bath

For 3 and more exposure modules

High Performance Water Bath - a perfect solution for heating 3 or more exposure modules.

VITROCELL® cultivation and exposure modules can be operated without an incubator. The modules can be heated to maintain a media temperature of 37º C (99º F). For the VITROCELL® 6/6, 12/12, 24/24 and 24/48 systems, heating is controlled electrically. All other modules are heated by warm water, either by water circuit or by a heating plate for optimal heat transferal to the module. This water bath is perfectly suitable for 3 or more exposure modules.

Ensuring accuracy in temperature control with a modern state-of-the-art design. These units provide best-in-class economical performance while remaining user-friendly by reducing unnecessary functions and focusing on reliability. Engineered from highest-quality components and materials, such as stainless steel for the baths, durable heating thermostats with a long operation life are guaranteed.

high-performance-water-bath-up.png

Features

  • For more than 3 modules

  • 5.7“ touchscreen and comfortable menu navigation

  • 5-point calibration

  • Temperature Range 25° C to 100° C

  • Prolonged operating life

  • Dimensions: 147 x 307 x 330 mm (W x D x H)

  • Delivered with cover plate and tubing connectors

  • Max. flow 27 l/min

  • Max. pump pressure 0,7 bar

  • Bath volume max. 6 Liters

  • Filling volume min. 4.9 Liters

VITROCELL® Vacuum Pump

Safe Flow Generation

VITROCELL® vacuum diaphragm pumps have been developed and manufactured to be chemically resistant. Thus, they are suitable for aggressive or corrosive gases and vapors alike.

The dry-running devices are specifically developed for laboratory applications and are maintenance-free. The pump exhaust line must be guided to a laboratory hood for proper extraction.

Vacuum-pump-up.png

Features

  • Compatible with all vapors and condensations

  • Chemically-resistant

  • Suitable for highly aggressive or corrosive gases

  • Maintenance-free

VITROCELL® Precision Pump System

Time-based supply of smallest liquid amounts

The VITROCELL® Precision Pump System was specifically developed and engineered for the time-based dosing of miniscule quantities (5-180 µl).

This stepper motor controlled diaphragm dosing pump system has been specially designed to accurately dose small quantities of liquid over long periods of time. It has two distinctive features - its scalability and durability (pump and electronics).

The VITROCELL® CF modules feature a medium exchange option.

precision-pump-system-overview.webp

Advantages

  • Simple and precise pump control (interval and volume)

  • Flexibility of the control process

  • Modular concept: various pump controllers can be interconnected

Features

  • Excellent repeatability

  • High stability in the long term

  • High chemical resistance

  • Fast priming and purging mode

  • Self priming

  • Run-dry ability

  • Prolonged operating life

  • Maintenance-free (>10‘000 h)

Technical data

  • Dispense mode: 5 - 180 µl

  • Priming mode (max.): 20 ml/min

  • Max. suction height: 4 mWg

  • Max. back pressure: 6 bar

  • Accuracy: ±2%

  • Repeatability: ±1%

  • Approved ambient temperature: +5 - +40° C

  • Approved liquid temperature: +5 - +80° C

  • Max. liquid viscosity: 150 cST

  • Dimensions Power Supply and Pump Controller: 240 x 200 x 88 mm (W x D x H)

  • Voltage: 100-240 V, 50/60 Hz

VITROCELL®EPITHELIX Nanopress

For application of powders to 3D cells

An efficient tool for the powder dosing of 3-D cell cultures.

The VITROCELL® EPITHELIX Nanopress enables particle and powder dosage in a precise and reproducible manner, thus making it an optimal tool for exposure to 3-D cell cultures.

The process was developed by EPITHELIX.

First, the test substance is mixed with Dextran powder. The mixing ratio can be varied according to the required dose/response relationship. Using the nanopress, tablets are produced so that they can be placed directly on the cell cultures. After incubation of approximately 15 min. the Dextran tablet dissolves and the cells are directly exposed to the test substance.

epithelix-nanopress-up.png

How the EPITHELIX Nanopress works

Epithelix-Pic1-IMG0328-640x427.webp

1. Add Dextran powder to the Nanopress.

2. Fill the reservoir.

3. Remove surplus powder from the press surface with a cell scraper.

Epithelix-Pic6-IMG-0337-640x427.webp

4. Push the lever down.

5. Push the button to expel the tablet.

6. Remove the tablet.

Epithelix-Pic7-IMG-0345-640x427.webp
Epithelix-Pic8-IMG-0349-640x427.webp
Epithelix-Pic9-IMG-0351-640x427.webp

7. Place the tablet on the apical surface.

8. Incubate for approx. 15 minutes.

9. The Dextran tablet is entirely dissolved.

Features

  • Exposure to a defined and small dose possible

  • Easy to use

  • Perfect screening tool

VITROCELL® Leak Tester

For easy and efficient leak detection

Regular leak tests are mandatory in aerosol research. A leak caused by a forgotten connection or defective o-ring may have a significant influence of the aerosol exposure process.

We recommend to carry out a leak test prior to the experiment and as part of the celaning or service routine.

The respective leakage rate is expressed in mm/60 s of fluid drop in the test column. If the instrument is connected to a closed system, this rate determines the amount of air entering the system from the outside as induces by a leak.

Leak-Tester.jpg.webp.jpeg

Features

  • Easy and sensitive method

  • Applicable for all aerosol ducts and exposure modules

  • Precise setting on setpoint mark due to bellow with venting valve

  • Solid pod for optimal positioning